Influence of time delay in calibration systems for instrument transformers with digital output
Author
Juvik, J.I.
Author_Institution
SP Swedish Nat. Testing & Res. Inst., Boras, Sweden
fYear
2000
fDate
14-19 May 2000
Firstpage
359
Lastpage
360
Abstract
Calibration of instrument transformers with digital output requires determination of the time delay in the digital system. The time delay of reference systems must therefore be accurately known. Measurements on SP´s measurement system have shown the time delay to be 190/spl plusmn/21 ns. This result is very promising and adequate for the system.
Keywords
calibration; delays; instrument transformers; calibration systems; digital output; instrument transformers; reference systems; time delay; Calibration; Delay effects; Displacement measurement; Frequency measurement; Instrument transformers; Phase measurement; Pulse measurements; Sampling methods; Software measurement; Time measurement;