DocumentCode :
2187001
Title :
Improved interstitial and intravascular MR imaging coils
Author :
Konyer, N.B. ; Viswanathan, R. ; Raghavan, R. ; Mills, G. ; Brady, M. ; Bronskill, M.J.
Author_Institution :
Sunnybrook & Women´´s Coll. Health Sci. Centre, Toronto Univ., Ont., Canada
fYear :
2002
fDate :
2002
Firstpage :
963
Lastpage :
965
Abstract :
An improved intravascular (IV) and interstitial (IS) prototype magnetic resonance (MR) imaging coil design offers dramatic gains in signal-to-noise ratio (SNR) and imaging volume. A multi-turn rectangular solenoid was wound inside catheters with 1.7 mm (IV) and 3.0 mm (IS) diameters. A factor of 200 in peak SNR gain over the standard birdcage head coil was observed, with the region of improved SNR persisting for a radial distance of 10 and 15 mm respectively. Very high resolution ex-vivo intravascular images and very high sensitivity MR spectroscopy (MRS) are demonstrated. This coil design offers interesting performance options in SNR and imaging volume compared to existing FDA-approved catheter coil designs.
Keywords :
NMR spectroscopy; biomedical MRI; blood vessels; catheters; solenoids; 1.7 mm; 10 mm; 15 mm; 3.0 mm; catheters; coil design; imaging volume; interstitial MR imaging coils; intravascular MR imaging coils; magnetic resonance imaging coil design; multi-turn rectangular solenoid; peak SNR gain; radial distance; signal-to-noise ratio; very high resolution ex-vivo intravascular images; very high sensitivity MR spectroscopy; Catheters; Coils; Head; Magnetic resonance; Magnetic resonance imaging; Prototypes; Signal design; Signal to noise ratio; Solenoids; Wounds;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Imaging, 2002. Proceedings. 2002 IEEE International Symposium on
Print_ISBN :
0-7803-7584-X
Type :
conf
DOI :
10.1109/ISBI.2002.1029422
Filename :
1029422
Link To Document :
بازگشت