    Performance test and application of GSO detector assembly for a continuous blood sampling system

  • Author

    Kudomi, Nobuyuki ; Choi, Eunjoo ; Yamamoto, Seiichi ; Watabe, Hiroshi ; Kim, Kyeong-Min ; Hayashi, Takuya ; Ogawa, Mikako ; Teramoto, Noboru ; Sakamoto, Eishi ; Iida, Hidehiro

  • Author_Institution
    Nat. Cardiovascular Center Res. Inst., Osaka, Japan
  • Volume
    3
  • fYear
    2002
  • fDate
    10-16 Nov. 2002
  • Firstpage
    1648
  • Abstract
    A new input function monitoring system has been developed using a GSO detector for both PET and SPECT quantitative studies. The paired assembly of crystals provided an absolute sensitivity of approximately 7% for PET tracers and 70% for 99mTc and 201Tl (SPECT tracers). This system was applied to clinical use and animal study such as monkey and rat. This study demonstrates that the present system can he of use in both clinical and small animal studies using SPECT and PET tracers.
  • Keywords
    haemodynamics; positron emission tomography; single photon emission computed tomography; solid scintillation detectors; 201Tl; 99mTc; GSO detector; Gd-Si-O; PET; SPECT; Tc; Tl; blood sampling system; monkey; rat; sensitivity; Animals; Assembly systems; Blood; Detectors; Monitoring; Photonic crystals; Positron emission tomography; Sampling methods; Single photon emission computed tomography; System testing;