    Development of a calibration system of digital recorders for impulse tests

  • Author

    Jiawei, Liu Na Chen

  • Author_Institution
    Electr. Magn. Div., Nat. Inst. of Metrol., Beijing
  • fYear
    2008
  • fDate
    8-13 June 2008
  • Firstpage
    250
  • Lastpage
    251
  • Abstract
    The configuration of a calibration system of digital recorder (digitizer) for HV impulse measurement is described. The system is developed based on Labview. Its calibration methods are according to IEC61083-1-2001, excepted for a fitting algorithm introduced in the determination of static non-linearity for digital recorder. The calibration system has been applied to the calibration of 8-bit and 300 MHz digital recorder.
  • Keywords
    IEC standards; analogue-digital conversion; calibration; impulse testing; recorders; virtual instrumentation; HV impulse measurement; IEC61083-1-2001; Labview; calibration system; digital recorder; digitizer; impulse test; static nonlinearity; Calibration; Dynamic voltage scaling; Impulse testing; Interference; Noise level; Nonlinear dynamical systems; Oscilloscopes; Pulse measurements; System testing; Voltmeters;