• DocumentCode
    2284558
  • Title

    High accuracy evaluation of CENAM´s primary frequency standard: second order Doppler shift

  • Author

    Lopez, Sebastian ; Lopez, M. ; Dominguez, I. ; de Carlos, E.

  • fYear
    2002
  • fDate
    16-21 June 2002
  • Firstpage
    454
  • Lastpage
    455
  • Abstract
    The Centro Nacional de Metrologia in Mexico has developed an optically pumped cesium beam frequency standard, the CsOP-1. Currently, the major sources of systematic effects are being analysed and their corresponding frequency shift effects in the CsOp-1. In this work, we re-evaluate the second order Doppler shift by taking into account the modulation parameter.
  • Keywords
    Doppler shift; caesium; frequency standards; measurement uncertainty; modulation; optical pumping; Cs; frequency shift effects; measurement uncertainty; modulation parameter; optically pumped cesium beam frequency standard; second order Doppler shift evaluation; systematic accuracy effect analysis; Amplitude modulation; Atom optics; Doppler shift; Frequency; Laser excitation; Microwave ovens; Optical beams; Optical modulation; Optical pumping; Standards development;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
  • Conference_Location
    Ottawa, Ontario, Canada
  • Print_ISBN
    0-7803-7242-5
  • Type

    conf

  • DOI
    10.1109/CPEM.2002.1034918
  • Filename
    1034918