• DocumentCode
    229459
  • Title

    Evaluation and validation of a national WR-15 (50 to 75 GHz) power measurement system

  • Author

    Xiaohai Cui ; Yu Song Meng ; Yueyan Shan ; Wenze Yuan ; Chao Ma ; Yong Li

  • Author_Institution
    Nat. Inst. of Metrol., Beijing, China
  • fYear
    2014
  • fDate
    4-5 Dec. 2014
  • Firstpage
    1
  • Lastpage
    3
  • Abstract
    In this paper, we present some preliminary results from evaluation and validation of a national WR-15 (50 to 75 GHz) power measurement system, which has been upgraded for thermistor sensor (thermistor mount) measurements recently. The system is a physical realization of direct comparison transfer technique for microwave power sensor calibration. It has been evaluated and validated referring to a WR-15 microcalorimeter measurement system. Good performance has been observed preliminary, and some necessary improvement works have been planned in near future.
  • Keywords
    calibration; calorimeters; measurement systems; microwave detectors; microwave measurement; power measurement; temperature sensors; thermistors; National WR-15 power measurement system; WR-15 microcalorimeter measurement system; direct comparison transfer technique; frequency 50 GHz to 75 GHz; microwave power sensor calibration; thermistor mount; thermistor sensor; Calibration; Microwave measurement; NIST; Power measurement; Thermistors; Uncertainty; Direct comparison transfer; microcalorimeter; power measurement; thermistor mount;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Measurement Conference (ARFTG), 2014 84th ARFTG
  • Conference_Location
    Boulder, CO
  • Type

    conf

  • DOI
    10.1109/ARFTG.2014.7013418
  • Filename
    7013418