Comparison Between Thermoelectric and Bolometric Microwave Power Standards
Author
Brunetti, L. ; Oberto, L. ; Sellone, M. ; Vremera, E.T.
Author_Institution
Electromagn. Div., Ist. Naz. di Ricerca Metrol., Turin, Italy
Volume
62
Issue
6
fYear
2013
fDate
Jun-13
Firstpage
1710
Lastpage
1715
Abstract
This paper describes a comparison of microwave power standards based on thermoelectric sensors to an analogous standard based on bolometric sensors. Measurements have been carried out with the classical twin-type microcalorimeter, fitted with N-connector test ports suitable for the frequency band 0.05-18 GHz. An appropriate measurand definition is given as being suitable for both standard types. A system accuracy assessment is performed applying the Gaussian error propagation through the mathematical models that interpret the microcalorimeter response in each case. The results highlight advantages and weaknesses of each power standard type.
Keywords
Gaussian noise; bolometers; calorimeters; measurement errors; measurement standards; microsensors; microwave measurement; temperature sensors; thermoelectric devices; Gaussian error propagation; N-connector test port; analogous standard; bolometric sensor; frequency 0.05 GHz to 18 GHz; mathematical model; microwave power standard; system accuracy assessment; thermoelectric sensor; twin type microcalorimeter; Broadband microcalorimeter; microwave measurements; microwave standards; power measurement; thermistors; thermoelectric devices;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2013.2253914
Filename
6502711
Link To Document