DocumentCode :
2356604
Title :
Novel method of automatic auscultation for blood pressure measurement using pulses in cuff pressure and korotkoff sound
Author :
Park, Dk ; Oh, HS ; Kang, JH ; Kim, IY ; Chee, YJ ; Lee, JS
Author_Institution :
Hanyang Univ., Seoul
fYear :
2008
fDate :
14-17 Sept. 2008
Firstpage :
181
Lastpage :
184
Abstract :
In the NIBP (Non-Invasive Blood Pressure) measurement, we need well trained observer and there are differences in reading values between observers. We suggest new automated auscultation algorithms to select the proper epoch of K-sound for each heart beat. The sound for each epoch is converted into frequency domain and total power was calculated. We decided SBP(Systolic Blood Pressure) point when there was abrupt jump from the baseline. DBP (Diastolic blood pressure) was decided when there was the biggest drop in total power of successive beats. For the validation of our suggested method, human observer assessed SBP/DBP according to EHS (European Hypertension Society) recommendation. The mean differences between the automatic method and traditional auscultation method were 2.0 mmHg for SBP and 3.1 mmHg for DBP. We also took the paired t-test with the two results, and there was no statistically significant difference between these two methods.
Keywords :
acoustic measurement; blood pressure measurement; cardiovascular system; medical signal processing; European Hypertension Society; K-sound; automated auscultation algorithm; cuff pressure; diastolic blood pressure; korotkoff sound; noninvasive blood pressure measurement; paired t-test; systolic blood pressure; Arteries; Band pass filters; Blood pressure; Frequency domain analysis; Heart beat; Humans; Microphones; Pressure measurement; Pulse measurements; Timing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology, 2008
Conference_Location :
Bologna
ISSN :
0276-6547
Print_ISBN :
978-1-4244-3706-1
Type :
conf
DOI :
10.1109/CIC.2008.4749007
Filename :
4749007
Link To Document :