DocumentCode
236192
Title
Evaluation of linearity in digital voltmeters using PJVS
Author
Maruyama, Mihoko ; Katayama, Kengo ; Yonezawa, Teru ; Kanai, Teruto ; Yoshida, Hiroyuki ; Iwasa, Akio ; Urano, Chiharu ; Kaneko, Nobu-hisa
Author_Institution
Nat. Metrol. Inst. of Japan (NMIJ), AIST, Tsukuba, Japan
fYear
2014
fDate
24-29 Aug. 2014
Firstpage
262
Lastpage
263
Abstract
We are evaluating the linearity characteristics of digital voltmeters (DVMs) using a programmable Josephson voltage standard (PJVS) system. The PJVS system is composed of a 524 288 junction array and a 10 K compact cooler, which enables to generate an arbitrary output voltage of more than 10 V. The voltage signal output with the PJVS system was applied to DVMs, and the linearity characteristics were investigated based on the reading values of the DVMs. As a result, we found that the linearity characteristics can be evaluated with the expanded uncertainty of 0.1 μV/V or better for the 10 V or smaller ranges.
Keywords
Josephson effect; digital voltmeters; measurement standards; voltage measurement; DVM; PJVS system; arbitrary output voltage generation; compact cooler; digital voltmeter; junction array; programmable Josephson voltage standard system; temperature 10 K; voltage 10 V; voltage signal output; Accuracy; Arrays; Linearity; Standards; Switches; Uncertainty; Voltage measurement; DC voltage; DVM; Josephson junction array; PJVS; linearity; mechanical cooler; precision measurements;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location
Rio de Janeiro
ISSN
0589-1485
Print_ISBN
978-1-4799-5205-2
Type
conf
DOI
10.1109/CPEM.2014.6898359
Filename
6898359
Link To Document