DocumentCode
236648
Title
Improvements on INRIM coaxial microcalorimeter
Author
Brunetti, L. ; Oberto, L. ; Sellone, M. ; Shoaib, Nosherwan ; Vremera, E.
Author_Institution
Ist. Naz. di Ricerca Metrol. (INRIM), Turin, Italy
fYear
2014
fDate
24-29 Aug. 2014
Firstpage
750
Lastpage
751
Abstract
This paper describes hardware and software improvements of the INRIM coaxial microcalorimeter together with their effect on the primary power standard realization in the frequency band 0.05 - 40 GHz. A better temperature and power stabilization turned out to provide an improved signal/noise ratio and a drift reduction in every working condition of the microcalorimeter. Results are presented in form of a 2.92 mm thermoelectric power sensor calibration together with graphs that show the improved stability and repeatability of the measurement system.
Keywords
UHF measurement; calibration; calorimeters; microsensors; microwave measurement; millimetre wave measurement; power measurement; thermoelectric devices; INRIM coaxial microcalorimeter; drift reduction; frequency 0.05 GHz to 40 GHz; power stabilization; primary power standard; signal-noise ratio; size 2.92 mm; temperature stabilization; thermoelectric power sensor calibration; Calibration; Microwave measurement; Microwave theory and techniques; Power measurement; Standards; Temperature measurement; Thermal stability; Broadband microcalorimeter; microwave measurements; microwave standards; power measurement; thermoelectric devices;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location
Rio de Janeiro
ISSN
0589-1485
Print_ISBN
978-1-4799-5205-2
Type
conf
DOI
10.1109/CPEM.2014.6898603
Filename
6898603
Link To Document