• DocumentCode
    2367553
  • Title

    Multi-frequency interstitial ultrasound applicator for conformal thermal therapy

  • Author

    Chopra, Rajiv ; Luginbuhl, Chris ; Weymouth, Alfred J. ; Foster, F. Stuart ; Bronskill, Michael J.

  • Author_Institution
    Dept. of Med. Biophys., Toronto Univ., Ont., Canada
  • Volume
    2
  • fYear
    2001
  • fDate
    2001
  • Firstpage
    1335
  • Abstract
    The ability to deliver conformal thermal therapy was investigated for multi-frequency interstitial heating applicators incorporating planar transducers and device rotation. Thermal lesions were generated in excised tissue and the influence of exposure parameters such as frequency, power, and device rotation was studied. Lesion radius was dependent on all exposure parameters studied, and a maximum radius of 18.0±0.6 mm was achieved. This applicator design and delivery strategy make conformal interstitial thermal therapy possible
  • Keywords
    biomedical ultrasonics; biothermics; radiation therapy; tumours; ultrasonic transducers; 17.4 to 18.6 mm; conformal thermal therapy; device rotation; exposure parameters; lesion radius; multi-frequency interstitial ultrasound applicator; planar transducers; thermal lesions; tumours; Acoustic transducers; Applicators; Force measurement; Frequency; Heating; Lesions; Medical treatment; Power generation; Ultrasonic imaging; Ultrasonic transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2001 IEEE
  • Conference_Location
    Atlanta, GA
  • Print_ISBN
    0-7803-7177-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2001.991967
  • Filename
    991967