• DocumentCode
    2392647
  • Title

    Performance comments on manganin resistance gauges under high static pressures

  • Author

    Tian-You, Yuan ; Yan-Chong, Lu ; Bao-Hui, Yuan

  • Author_Institution
    Xi Modern Chem. Inst., China
  • fYear
    1989
  • fDate
    25-27 Apr 1989
  • Firstpage
    559
  • Lastpage
    562
  • Abstract
    The performance of manganin resistance gauges under high static pressure is evaluated. The accuracy of several interpolation- (or extrapolation-) based mathematical-models is compared. It is shown that if the gauges used at pressures below 1 GPa are calibrated at a given constant temperature and used at that temperature, the complete mathematical-model is unnecessary. However, a complete model for the variation of resistance with both pressure and temperature is required in the pressure range from 1 GPa up to 20 GPa, because the temperature changes due to compression or expansion of the pressure-transmitting medium can be quite pronounced
  • Keywords
    calibration; copper alloys; extrapolation; high-pressure techniques; interpolation; manganese alloys; nickel alloys; pressure transducers; 1 GPa; 1 GPa to 20 GPa; Cu-Mn-Ni; calibration; compression; expansion; extrapolation; high pressure techniques; high static pressures; interpolation; manganin resistance gauges; mathematical-models; temperature changes; Chemistry; Coils; Electrical resistance measurement; Extrapolation; Manufacturing; Pressure measurement; Reproducibility of results; Stability; Wire; Wounds;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1989. IMTC-89. Conference Record., 6th IEEE
  • Conference_Location
    Washington, DC
  • Type

    conf

  • DOI
    10.1109/IMTC.1989.36923
  • Filename
    36923