DocumentCode
2392647
Title
Performance comments on manganin resistance gauges under high static pressures
Author
Tian-You, Yuan ; Yan-Chong, Lu ; Bao-Hui, Yuan
Author_Institution
Xi Modern Chem. Inst., China
fYear
1989
fDate
25-27 Apr 1989
Firstpage
559
Lastpage
562
Abstract
The performance of manganin resistance gauges under high static pressure is evaluated. The accuracy of several interpolation- (or extrapolation-) based mathematical-models is compared. It is shown that if the gauges used at pressures below 1 GPa are calibrated at a given constant temperature and used at that temperature, the complete mathematical-model is unnecessary. However, a complete model for the variation of resistance with both pressure and temperature is required in the pressure range from 1 GPa up to 20 GPa, because the temperature changes due to compression or expansion of the pressure-transmitting medium can be quite pronounced
Keywords
calibration; copper alloys; extrapolation; high-pressure techniques; interpolation; manganese alloys; nickel alloys; pressure transducers; 1 GPa; 1 GPa to 20 GPa; Cu-Mn-Ni; calibration; compression; expansion; extrapolation; high pressure techniques; high static pressures; interpolation; manganin resistance gauges; mathematical-models; temperature changes; Chemistry; Coils; Electrical resistance measurement; Extrapolation; Manufacturing; Pressure measurement; Reproducibility of results; Stability; Wire; Wounds;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference, 1989. IMTC-89. Conference Record., 6th IEEE
Conference_Location
Washington, DC
Type
conf
DOI
10.1109/IMTC.1989.36923
Filename
36923
Link To Document