DocumentCode :
2412495
Title :
Detection and prediction of concentrations of neurotransmitters using voltammetry and pattern recognition
Author :
Sazonova, Nadezhda ; Njagi, John I. ; Marchese, Zachary S. ; Ball, Michael S. ; Andreescu, Silvana ; Schuckers, S.
Author_Institution :
Dept. of Electr. & Comput. Eng., Clarkson Univ., Potsdam, NY, USA
fYear :
2009
fDate :
3-6 Sept. 2009
Firstpage :
3493
Lastpage :
3496
Abstract :
Neurotransmitters (NTs) are substances in the brain which are responsible for the transmission of neurological impulses. Changes in their concentrations are associated with numerous behavioral and physiological processes and neurological disorders. As opposed to the traditional chromatographic and capillary electrophoresis, using electrochemical sensors is a fast and inexpensive way to determine concentrations of NTs. In this study we measure the combination of dopamine (DA) and serotonin (SE) with glassy carbon electrodes and differential pulse voltammetry. The major challenge using this method is to differentiate between different NTs, since the signal obtained from the electrode represents the interactive effect of both NTs present. We address this problem through methods of pattern recognition which relate the voltammetric measurements provided by the sensor to the concentration of individual NTs. Two methods of pattern recognition were applied (PCR and PLS-regression). The best rates of correct classification for the validation sets ranged at 42-62% (DA) and 33-50% (SE). When the ranges for correct prediction were extended to include one level above and below the true concentration level, the rates values ranged at 81-91% (DA) and 91-100%(SE). These findings suggest that pattern recognition can be used to model the interaction between different neurotransmitters to predict actual concentrations of neurotransmitters using voltammetry.
Keywords :
biomedical measurement; brain; electrochemical sensors; least squares approximations; molecular biophysics; neurophysiology; pattern recognition; principal component analysis; proteins; regression analysis; voltammetry (chemical analysis); brain; differential pulse voltammetry; dopamine; electrochemical sensors; glassy carbon electrodes; neurological impulse transmission; neurotransmitters; partial least squares regression; pattern recognition; principal component regression; serotonin; Computer Simulation; Dopamine; Electrochemistry; Electrodes; Electrophoresis, Capillary; Humans; Least-Squares Analysis; Microelectrodes; Nervous System Diseases; Neurotransmitter Agents; Pattern Recognition, Automated; Polymerase Chain Reaction; Principal Component Analysis; Regression Analysis; Serotonin;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
Conference_Location :
Minneapolis, MN
ISSN :
1557-170X
Print_ISBN :
978-1-4244-3296-7
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/IEMBS.2009.5334575
Filename :
5334575
Link To Document :
بازگشت