3D Radiation Detectors: Charge Collection Characterisation and Applicability of Technology for Microdosimetry
Author
Tran, Linh T. ; Prokopovich, Dale A. ; Petasecca, Marco ; Lerch, Michael L. F. ; Kok, Angela ; Summanwar, Anand ; Hansen, Thor-Erik ; Da Via, C. ; Reinhard, Mark I. ; Rosenfeld, Anatoly B.
Author_Institution
Centre for Med. Radiat. Phys., Univ. of Wollongong, Wollongong, NSW, Australia
Volume
61
Issue
4
fYear
2014
fDate
Aug. 2014
Firstpage
1537
Lastpage
1543
Abstract
A study of charge collection in SINTEF 3D active edge silicon detectors was carried out at ANSTO using Ion Beam Induced Charge (IBIC) technique. An IBIC study has shown that several different geometries of 3D detectors have full depletion under low applied bias. The effect of fast neutron and gamma radiation on their charge collection efficiency was also investigated. A 3D active edge silicon detector technology has demonstrated extremely promising performance for application of the 3D Sensitive Volumes (SVs) fabrication methods to SOI microdosimetry.
Keywords
dosimetry; silicon radiation detectors; 3D Sensitive Volumes fabrication methods; 3D radiation detectors; IBIC study; IBIC technique; Ion Beam Induced Charge; SINTEF 3D active edge silicon detectors; SOI microdosimetry; charge collection characterisation; fast neutron effect; gamma radiation effect; microdosimetry; Detectors; Electrodes; Fabrication; Image edge detection; Neutrons; Silicon; Three-dimensional displays; 3D detector; IBIC; charge collection; microdosimetry; radiation damage;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2014.2301729
Filename
6759760
Link To Document