DocumentCode :
2459800
Title :
Performance evaluation of a microcalorimeter
Author :
Jeong Il Park ; Jeong Hwan Kim
Author_Institution :
Korea Res. Inst. of Stand. & Sci., Taejon, South Korea
fYear :
1996
fDate :
17-21 June 1996
Firstpage :
458
Lastpage :
459
Abstract :
A waveguide microcalorimeter in the frequency range of 18 GHz-26.5 GHz is described. The systematic uncertainties due to microcalorimeter correction factor are discussed. Microcalorimeter using a curve fitting method can reduce the measurement time of the effective efficiencies measurement by a factor of 2. The results of measurements are presented and compared with the data from NIST.
Keywords :
calorimeters; curve fitting; measurement errors; measurement standards; microwave measurement; power measurement; 18 to 26.5 GHz; NIST; correction factor; curve fitting method; effective efficiencies measurement; measurement time; microwave power measurement; systematic uncertainties; waveguide microcalorimeter; Bolometers; Circuits; Curve fitting; Microwave measurements; Power measurement; Power transmission lines; Radio frequency; Steady-state; Thermistors; Time measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 1996 Conference on
Conference_Location :
Braunschweig, Germany
Print_ISBN :
0-7803-3376-4
Type :
conf
DOI :
10.1109/CPEM.1996.547164
Filename :
547164
Link To Document :
بازگشت