Title :
P5B-2 Imaging Microbubble Destruction/Replenishment with Nakagami Distribution
Author :
Tsui, Po-Hsiang ; Li, Meng-Lin ; Chang, Chien-Cheng ; Yeh, Chih-Kuang
Author_Institution :
Acad. Sinica, Taipei
Abstract :
Contrast-assisted techniques that quantitatively estimate time-intensity curves (TICs) within the perfusion area have been applied in many disease models. In this study, we monitored the microbubble replenishment by a proposed new approach called the time-Nakagami-parameter curve (TNC) obtained from Nakagami-parameter images for quantifying the microvascular flow velocity. The feasibility of using the TNC to estimate the microvascular flow was evaluated in phantom experiments of the TIC and TNC. The rates of increase of the TIC and TNC were expressed by the rate constants beta1 and betaN of a monoexponential model, respectively. The results showed that betaN quantifies the flow velocity similarly to the conventional beta1. In particular, the TNC-based technique conditionally tolerates the presence of nonperfused areas (e.g., surrounding tissues or vessel walls) in the region of interest without requiring application of an additional wall filter to cancel the influences of clutter echoes on the flow estimation.
Keywords :
biomedical ultrasonics; blood vessels; bubbles; haemodynamics; haemorheology; ultrasonic imaging; Nakagami distribution; biological tissues; clutter echoes; contrast-assisted techniques; microbubble destruction; microbubble replenishment; microvascular flow velocity; perfusion area; phantoms; time-Nakagami-parameter curve; time-intensity curve; ultrasonic imaging; vessel walls; Backscatter; Blood flow; Fluid flow measurement; High-resolution imaging; Information filtering; Information filters; Nakagami distribution; Statistical distributions; Ultrasonic imaging; Ultrasonic variables measurement;
Conference_Titel :
Ultrasonics Symposium, 2007. IEEE
Conference_Location :
New York, NY
Print_ISBN :
978-1-4244-1384-3
Electronic_ISBN :
1051-0117
DOI :
10.1109/ULTSYM.2007.551