Abstract :
The authors evaluated an algorithm present in the DDD pacemaker Chorus 6003 (Ela Medical) able to detect and treat the pacemaker-mediated tachycardia (PMT) and able to reprogram automatically some parameters to prevent new episodes of tachycardia. Ten patients were implanted with Chorus 6003 and submitted to 24-hour Holter monitoring to evaluate the algorithm performance. PMTs were revealed in 7/10 patients, and a total of 261 episodes were detected. All episodes were terminated, although 71 episodes were interrupted during detection and confirmation phases. In 3 patients closely coupled PMT episodes activated the autoreprogram function. In the study, the algorithm showed a high sensitivity and specificity
Keywords :
medical computing; pacemakers; 24 hr; DDD pacemaker Chorus 6003; Holter monitoring; algorithm performance; automatic algorithm; automatic reprogramming; autoreprogram function; pacemaker mediated tachycardia; self-reprogramming pacemakers; Biomedical monitoring; Delay; Heart rate variability; Pacemakers; Phase detection; Physiology; Protection; Rhythm; Sensitivity and specificity; Stress;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology 1991, Proceedings.
Conference_Location :
Venice
Print_ISBN :
0-8186-2485-X
Type :
conf
DOI :
10.1109/CIC.1991.169069
Filename :
169069
Link To Document :
بازگشت