Title :
Accuracy of RF-vector-voltmeters
Author :
Gärtner, Uwe F. ; Schiek, Burkhard
Author_Institution :
Inst. fuer Hoch- und Hoechstfrequenztech., Ruhr-Univ. Bochum, West Germany
Abstract :
A general analysis of quadrature and crosstalk errors in vector-voltmeter circuits is presented. The two error sources can be separated and quantitatively determined if the periodic phase-error function of the detector is available. Several schemes for determining the phase-error function of a vector-voltmeter are discussed. A self-calibrating error-detection principle is proposed which needs no reference standard and eliminates mismatch error effects
Keywords :
calibration; crosstalk; measurement errors; voltmeters; RF-vector-voltmeters; crosstalk errors; error sources; phase-error function; self-calibrating error-detection; vector-voltmeter circuits; Distortion; Gas detectors; Infrared detectors; Laboratories; Measurement errors; Microwave measurements; Phase detection; Phase measurement; Signal analysis; Signal detection;
Conference_Titel :
Instrumentation and Measurement Technology Conference, 1988. IMTC-88. Conference Record., 5th IEEE
Conference_Location :
San Diego, CA
DOI :
10.1109/IMTC.1988.10862