DocumentCode
2560477
Title
An adaptive method for automatic calibration of interpolation time interval digitizer with picosecond resolution
Author
Pelka, Ryszard
Author_Institution
Warsaw Acad. of Technol., Poland
fYear
1990
fDate
11-14 June 1990
Firstpage
15
Lastpage
16
Abstract
An adaptive method is presented for real-time calibration of an interpolation time-interval digitizer (TID) with picosecond resolution. Reduction of calibration uncertainty to the level of 3 ps (RMS) has been achieved. Experimental results and comparison with some existing methods for TID calibration are also presented. The adaptive method was developed to solve a calibration problem that emerged in the course of developing an advanced TID for a satellite laser ranging systems.<>
Keywords
adaptive systems; analogue-digital conversion; calibration; computerised instrumentation; interpolation; measurement by laser beam; time measurement; adaptive method; automatic calibration; interpolation time interval digitizer; picosecond resolution; real-time calibration; satellite laser ranging; time measurement; Analog-digital conversion; Calibration; Counting circuits; Error analysis; Interpolation; Laser theory; Measurement errors; Measurement uncertainty; Satellites; Time measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements, 1990. CPEM '90 Digest., Conference on
Conference_Location
Ottawa, Ontario, Canada
Type
conf
DOI
10.1109/CPEM.1990.109903
Filename
109903