DocumentCode
2563451
Title
Investigating the use of multimeters to measure quantized Hall resistance standards
Author
Cage, M.E. ; Yu, D.Y. ; Jeckelmann, B.M. ; Steiner, R.L. ; Duncan, R.V.
Author_Institution
Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
fYear
1990
fDate
11-14 June 1990
Firstpage
332
Lastpage
333
Abstract
A new generation of digital multimeters was used to compare the ratios of the resistances of wire-wound reference resistors and quantized Hall resistances (QHRs). Specifically, a digital multimeter was used to compare the DC voltage ratio of a QHR sample with that of a reference resistor connected in series with the sample. The accuracies are better than 0.1 p.p.m. for ratios as large as 4:1 if the multimeters are calibrated with a Josephson array.<>
Keywords
Josephson effect; calibration; digital multimeters; electric resistance measurement; measurement standards; quantum Hall effect; resistors; DC voltage ratio; Josephson array; digital multimeters; multimeters; quantized Hall resistance standards; reference resistor; wire-wound reference resistors; Calibration; Current measurement; Electric resistance; Electrical resistance measurement; Laboratories; Measurement standards; NIST; Power generation; Resistors; Voltage;