Title :
A new microcalorimeter in K/sub a/ band
Author :
Fang Yan ; Wu Youzhang
Author_Institution :
Nat. Inst. of Metrol., Beijing, China
Abstract :
Based on an analysis of a thermistor mount which is used as the load of a microcalorimeter, a novel type of K/sub a/-band microcalorimeter is reported. The total uncertainty of the microcalorimeter in calibrating directly the effective efficiency eta /sub e/ of commercial thermistor mounts has been estimated to be +or-0.2 to +or-0.5%.<>
Keywords :
calibration; calorimeters; microwave measurement; thermistors; K/sub a/ band; calorimeter; effective efficiency; load; microcalorimeter; thermistor mount; Automatic generation control; Bridge circuits; Feedback control; Metrology; Programmable control; Temperature control; Temperature measurement; Temperature sensors; Thermistors; Uncertainty;
Conference_Titel :
Precision Electromagnetic Measurements, 1990. CPEM '90 Digest., Conference on
Conference_Location :
Ottawa, Ontario, Canada
DOI :
10.1109/CPEM.1990.110076