• DocumentCode
    2564061
  • Title

    A new microcalorimeter in K/sub a/ band

  • Author

    Fang Yan ; Wu Youzhang

  • Author_Institution
    Nat. Inst. of Metrol., Beijing, China
  • fYear
    1990
  • fDate
    11-14 June 1990
  • Firstpage
    396
  • Lastpage
    397
  • Abstract
    Based on an analysis of a thermistor mount which is used as the load of a microcalorimeter, a novel type of K/sub a/-band microcalorimeter is reported. The total uncertainty of the microcalorimeter in calibrating directly the effective efficiency eta /sub e/ of commercial thermistor mounts has been estimated to be +or-0.2 to +or-0.5%.<>
  • Keywords
    calibration; calorimeters; microwave measurement; thermistors; K/sub a/ band; calorimeter; effective efficiency; load; microcalorimeter; thermistor mount; Automatic generation control; Bridge circuits; Feedback control; Metrology; Programmable control; Temperature control; Temperature measurement; Temperature sensors; Thermistors; Uncertainty;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements, 1990. CPEM '90 Digest., Conference on
  • Conference_Location
    Ottawa, Ontario, Canada
  • Type

    conf

  • DOI
    10.1109/CPEM.1990.110076
  • Filename
    110076