Title :
Feedthrough EMI filter with ground isolation for cardiac pacemakers and implantable cardioverter defibrillators
Author :
Stevenson, Robert A.
Author_Institution :
Maxwell Technol., Energy Products Inc., Carson City, NV, USA
fDate :
29 Oct-1 Nov 1998
Abstract :
This paper describes an Isolated Ground Capacitor (IGC) electromagnetic interference (EMI) feedthrough filter that generally makes cardiac pacemakers and implantable cardioverter defibrillators (ICDs) immune to the RF fields generated by cellular phones and other emitters. The IGC filter consists of a monolithic ceramic capacitor (MLC) in a feedthrough configuration which is directly mounted onto the hermetic terminal of the implantable medical device. The IGC MLC employs unique electrode plate geometry and manufacturing techniques that provide for isolation of pacemaker or ICD circuitry from the titanium shield (ground plane)
Keywords :
active filters; ceramic capacitors; defibrillators; electromagnetic compatibility; electromagnetic shielding; pacemakers; radiofrequency interference; EMC; RF field immunity; cardiac pacemakers; cellular phones; electrode plate geometry; feedthrough EMI filter; ground isolation; hermetic terminal; implantable cardioverter defibrillators; implantable medical device; insertion loss; isolated ground capacitor filter; monolithic ceramic capacitor; shielded housing; voltage divider; Bioceramics; Capacitors; Cardiology; Cellular phones; Electromagnetic interference; Filters; Immune system; Implantable biomedical devices; Pacemakers; Radio frequency;
Conference_Titel :
Engineering in Medicine and Biology Society, 1998. Proceedings of the 20th Annual International Conference of the IEEE
Conference_Location :
Hong Kong
Print_ISBN :
0-7803-5164-9
DOI :
10.1109/IEMBS.1998.746210