DocumentCode :
2615541
Title :
Study of scintillation crystal array parameters for an advanced PET scanner dedicated to breast cancer imaging
Author :
Vandenbroucke, A. ; Levin, C.S.
Author_Institution :
Department of Radiology, Stanford University, USA
fYear :
2008
fDate :
19-25 Oct. 2008
Firstpage :
4914
Lastpage :
4919
Abstract :
We present measurements and simulation to determine the optimal crystal array parameters for an advanced PET scanner dedicated to breast imaging. We compared LSO versus LYSO arrays, measured different crystal sizes and various surface finishes and array configurations. We conclude that an 8 × 8 array of 1 × 1 × 1 mm3 LYSO crystals with a specular (VM2000) foil on all sides of each individual crystal except the diffuse exit window gives the optimal results for our configuration.
Keywords :
Avalanche photodiodes; Breast cancer; Crystalline materials; Crystals; Histograms; Nuclear and plasma sciences; Positron emission tomography; Size measurement; Surface finishing; Surface treatment; LSO; LYSO; PET; PSAPD; scintillation;