Title :
Partial volume artifact reduction in computed tomography
Author :
Hsieh, Jiang
Author_Institution :
Appl. Sci. Lab., G.E. Medical Systems, Milwaukee, WI, USA
Volume :
3
fYear :
1996
fDate :
16-19 Sep 1996
Firstpage :
567
Abstract :
In the computed tomography (CT) head scans, dark streaks connecting the petrous bones are often observed. These streaks can lead to severely degraded image quality and render the image unusable for diagnosis. In this paper, we analyze one the main sources of the streaking artifact the nonlinear partial volume. We have developed an algorithm that effectively combats these artifacts. The algorithm estimates the degree of partial voluming by calculating the variations of the bony structures between adjacent slices. It then performs the correction based on a quadratic error model. Various experiments were conducted to demonstrate the effectiveness of the correction scheme
Keywords :
computerised tomography; diagnostic radiography; interference suppression; medical image processing; adjacent slices; algorithm; bony structures; computed tomography; correction scheme; dark streaks; degraded image quality; diagnosis; head scans; nonlinear partial volume; partial volume artifact reduction; petrous bones; quadratic error model; streaking artifact; Attenuation; Computed tomography; Error correction; Head; Image reconstruction; Laboratories; Object detection; X-ray detection; X-ray detectors; X-ray imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Image Processing, 1996. Proceedings., International Conference on
Conference_Location :
Lausanne
Print_ISBN :
0-7803-3259-8
Type :
conf
DOI :
10.1109/ICIP.1996.560558
Filename :
560558
Link To Document :
بازگشت