DocumentCode :
2624161
Title :
Reconstruction of incident X-ray spectra using the Medipix2 detector
Author :
Talla, Patrick Takoukam ; Firsching, Markus ; Michel, Thilo ; Durst, Jurgen ; Anton, Gisela
Author_Institution :
Erlangen Centre for Astroparticle Physics (ECAP), Friedrich-Alexander-Universitÿt, Erwin-Rommel-Strasse 1, 91058, Germany
fYear :
2008
fDate :
19-25 Oct. 2008
Firstpage :
2459
Lastpage :
2461
Abstract :
The single photon counting pixel detector Medipix2 has been used in a variety of applications. Because of its spectral capability, the detector could be used as spectrometer. However the measured spectrum is not the real incident spectrum as the Medipix suffers from charge sharing because of its small pixel size. This implies that the energy information is smeared out heavily. Knowing the measured spectrum and the responses to monoenergetic exposure we applied some deconvolution methods to determine the incoming spectrum.
Keywords :
Application specific integrated circuits; Deconvolution; Energy measurement; Extraterrestrial measurements; Integral equations; Nuclear and plasma sciences; Optoelectronic and photonic sensors; Physics; X-ray detection; X-ray detectors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location :
Dresden, Germany
ISSN :
1095-7863
Print_ISBN :
978-1-4244-2714-7
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2008.4774852
Filename :
4774852
Link To Document :
بازگشت