DocumentCode :
2639210
Title :
Tissue micromotion induced stress around brain implants
Author :
Muthuswamy, J. ; Saha, R. ; Gilletti, A.
Author_Institution :
Harrington Dept. of Bioeng., Arizona State Univ., Tempe, AZ, USA
fYear :
2005
fDate :
12-15 May 2005
Firstpage :
102
Lastpage :
103
Abstract :
The long-term consequences of tissue micromotion against stationary brain implants are poorly understood. Our aim here is to measure surface micromotion in the rodent somatosensory cortex and estimate mechanical stresses induced in the brain tissue due to micromotion against stationary implants. A differential variable reluctance transducer (DVRT) was used in adult rats to monitor micromotion normal to the somatosensory cortex surface. Using finite element models of the brain, we then estimated shear and normal stresses in the brain tissue in the vicinity of the brain implants. Surface micromotion was observed to be few tens of microns due to pressure changes during respiration and 2-4μm due to vascular pulsatility. Maximum shear stress values of up to 2.5-3.5 KPa were estimated near the tip of a 50μm diameter implant. Tissue micromotion on the surface of the somatosensory cortex can lead to significant shear and normal stress build-up in the brain tissue in the vicinity of cylindrical brain implants. The impact of the mechanical stress on brain tissue viability and function under chronic in-vivo conditions needs to be assessed in future studies.
Keywords :
biological tissues; biomechanics; biomedical transducers; brain; finite element analysis; motion measurement; neurophysiology; physiological models; pneumodynamics; prosthetics; somatosensory phenomena; 2.5 to 3.5 KPa; 50 micron; adult rat; brain implant; brain tissue micromotion; chronic in-vivo condition; differential variable reluctance transducer; finite element model; mechanical stress estimation; respiration; rodent somatosensory cortex; shear stress; surface micromotion measurement; vascular pulsatility; Biomedical monitoring; Brain modeling; Elasticity; Finite element methods; Implants; Microelectrodes; Rats; Stress; Transducers; Viscosity;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microtechnology in Medicine and Biology, 2005. 3rd IEEE/EMBS Special Topic Conference on
Print_ISBN :
0-7803-8711-2
Type :
conf
DOI :
10.1109/MMB.2005.1548395
Filename :
1548395
Link To Document :
بازگشت