Abstract
In freehand 3D Ultrasound (3D-US) scanning one needs to track a B-mode scanner across a subject of interest to construct the 3D image. In this paper we use a compact IMU (inertial measurement unit), as an alternative method that have certain advantages compared to other conventional methods. However IMUs accumulate translational errors thus we study if we can overcome them by low pass filtering and making assumptions on the nature of the introduced drift. We also briefly expound on alternative methods of obtaining translation by only using either the acceleration output or the orientation output. We then interpret the results of our experiments and expound we finally conclude with further improvements that can be made. We hope that this paper would be a good introduction to research using 6-DOF (degrees of freedom) MEMS IMUs that are becoming more accessible and have a wide range of applications.
Keywords
image reconstruction; low-pass filters; micromechanical devices; position measurement; ultrasonic applications; ultrasonic imaging; B-mode scanner; MEMS; degrees of freedom; freehand 3D ultrasound scanning; inertial measurement unit; low pass filtering; Acceleration; Accelerometers; Acoustic devices; Acoustic signal detection; Biomedical optical imaging; Low pass filters; Measurement units; Micromechanical devices; Systems engineering and theory; Ultrasonic imaging; 3D acceleration; 3D orientation; 6-DOF; IMU; MEMS;
fLanguage
English
Publisher
ieee
Conference_Titel
Applied Electromagnetics, 2007. APACE 2007. Asia-Pacific Conference on
Conference_Location
Melaka
Print_ISBN
978-1-4244-1434-5
Electronic_ISBN
978-1-4244-1435-2
Type
conf
DOI
10.1109/APACE.2007.4603861
Filename
4603861
Link To Document