DocumentCode :
2658611
Title :
Enhancing the frequency stability of a millimeter wave network analyzer with an add-on unit
Author :
Eskelinen, Pekka ; Säily, Jussi
Author_Institution :
Helsinki Univ. of Technol., Espoo, Finland
fYear :
2003
fDate :
4-8 May 2003
Firstpage :
521
Lastpage :
523
Abstract :
A proper arrangement frequency generation greatly enhances the stability of commercial millimeter wave instrumentation, particularly in network analyzers operating above 110 GHz. When distributed crystal clocks of varying uncertainty classes are replaced by a centralized rubidium master oscillator, the phase and frequency errors of the entire measuring system are reduced by a factor of ten. Galvanic ground loops and load pulling effects can be entirely avoided in these physically large installations by using optical fibers as a synchronization path. Millimeter wave spectrum characteristics remain unaffected because no additional phase locked loops or frequency multipliers are needed. The suggested modification is of great help in prolonged near-field scanning measurements, where a constant phase and carrier frequency have to be maintained even for a couple of days. Repeatability is extended up to several months through GPS-based frequency steering, which however must be disabled during millimeter wave activity in order to avoid phase jumps.
Keywords :
atomic clocks; crystal oscillators; frequency measurement; frequency stability; millimetre wave oscillators; millimetre wave spectra; network analysers; rubidium; 110 GHz; GPS based frequency steering; Rb; crystal clocks; frequency errors; frequency generation; frequency stability; galvanic ground loops; load pulling effects; millimeter wave activity; millimeter wave network analyzer; millimeter wave spectrum; near field scanning measurements; optical fibers; phase errors; rubidium master oscillator; synchronization path; Clocks; Frequency measurement; Frequency synchronization; Galvanizing; Instruments; Millimeter wave measurements; Optical fibers; Oscillators; Phase measurement; Stability analysis;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Frequency Control Symposium and PDA Exhibition Jointly with the 17th European Frequency and Time Forum, 2003. Proceedings of the 2003 IEEE International
ISSN :
1075-6787
Print_ISBN :
0-7803-7688-9
Type :
conf
DOI :
10.1109/FREQ.2003.1275145
Filename :
1275145