A novel device to simulate low respiratory static compliances of pre-term infants measuring expansion tidal volume
Author :
Sardellitti, I. ; Silvestri, S. ; Cappa, P.
Author_Institution :
Sch. of Biomed. Eng., Univ. Campus Bio-Medico, Rome, Italy
Volume :
4
fYear :
2003
fDate :
17-21 Sept. 2003
Firstpage :
3090
Abstract :
A prototype has been realized and metrologically characterized in order to be utilized to test infant pulmonary ventilators in in-vitro use. The adopted design for the proposed tester is capable to simulate and to regulate independently the characteristic values of respiratory system compliance and functional residual capacity of pre-term infants. Compliance values lower than 0.10 ml/cmH2O in a tidal volume range lower than 10 ml have been obtained. Functional residual capacity can be varied between 20 and 50 ml with a 3 ml resolution. Expansion tidal volume and alveolar pressure can be continuously monitored. The realized device is inexpensive and shows acceptable accuracy values associated to the simulated parameters. Thus, it can be considered a valuable tool in the estimation on objective basis of the actual performances of neonatal ventilators.
Keywords :
biomedical equipment; diseases; haemodynamics; lung; paediatrics; patient treatment; ventilation; H2O; alveolar pressure; in-vitro use; infant pulmonary ventilators; low respiratory static compliance; lung simulator; metrologically characteristics; neonatal ventilation; novel device; pre-term respiratory mechanics; Biomedical measurements; Cross layer design; Lungs; Mechanical variables measurement; Medical treatment; Pediatrics; Respiratory system; System testing; Ventilation; Volume measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2003. Proceedings of the 25th Annual International Conference of the IEEE
ISSN :
1094-687X
Print_ISBN :
0-7803-7789-3
Type :
conf
DOI :
10.1109/IEMBS.2003.1280795
Filename :
1280795
Link To Document :
بازگشت