Development of a 2.4-mm coaxial microcalorimeter for RF and microwave power standards at KRISS
Author :
Kwon, Jae-Yong ; Kang, Tae-Won ; Kim, Jeong-Hwan ; Crowley, Tom
Author_Institution :
Center for Electromagn. Wave, Korea Res. Inst. of Stand. & Sci. (KRISS), Daejeon, South Korea
fYear :
2012
fDate :
1-6 July 2012
Firstpage :
734
Lastpage :
735
Abstract :
This paper introduces a 2.4-mm coaxial microcalorimeter we have been developing together with thermoelectric reference standards. It will become a national power standard up to 50 GHz. Design of some core parts and implementation of the microcalorimeter are described. Some preliminary results and future plan for system evaluation will be briefly discussed.
Keywords :
calorimeters; microsensors; microwave measurement; power measurement; thermoelectric devices; KRISS; RF power standard; coaxial microcalorimeter; microwave power standard; size 2.4 mm; thermoelectric reference standard; Conductors; Connectors; Microwave technology; NIST; Radio frequency; Thermostats; Adiabatic lines; calibration; coaxial transmission lines; power measurement; thermoelectric devices;