DocumentCode :
2776633
Title :
Data reduction techniques to facilitate wireless and long term AEEG epilepsy monitoring
Author :
Casson, Alexander J. ; Rodriguez-Villegas, Esther
Author_Institution :
Dept. of Electr. & Electron. Eng., Imperial Coll. London
fYear :
2007
fDate :
2-5 May 2007
Firstpage :
298
Lastpage :
301
Abstract :
Wireless ambulatory EEG (AEEG) monitoring over long periods of time is currently infeasible due to battery limitations and the EEG analysis time required. A detailed comparison of methods for reducing the amount of AEEG data is presented. It is concluded that a discontinuous recording scheme can alleviate both of the above problems. Discontinuous monitoring introduces data interpretation and practical issues which are discussed. With suitable low power algorithm implementations and realistic system expectations such systems are deemed to be feasible.
Keywords :
data reduction; electroencephalography; medical signal processing; patient monitoring; AEEG data; data interpretation; data reduction; discontinuous monitoring; discontinuous recording scheme; wireless ambulatory EEG monitoring; Batteries; Electrodes; Electroencephalography; Epilepsy; Magnetic heads; Neural engineering; Patient monitoring; Sampling methods; USA Councils; Wires;