• DocumentCode
    2807475
  • Title

    Large scale phased array and system for treatment of deep seated tissue

  • Author

    Daum, D.R. ; Hynynen, K.

  • Author_Institution
    Brigham & Women´´s Hosp., Harvard Med. Sch., Boston, MA, USA
  • Volume
    2
  • fYear
    1998
  • fDate
    1998
  • Firstpage
    1559
  • Abstract
    A custom built 512 channel system was designed to drive therapeutic ultrasound phased arrays. The system was used to drive a newly constructed 256 element phased array in a series of MR guided in vivo porcine experiments to coagulate deep seated tissue in the thigh, kidney, and liver. Thermal coagulation lesions greater than 5 cm3 in thigh and greater than 2 cm3 in the liver were generated using a single 20 second sonication. The feasibility of MR guided phased array ultrasound surgery to coagulate in vivo liver tissue in a large animal model has been shown
  • Keywords
    biomedical equipment; biomedical ultrasonics; coagulation; kidney; liver; surgery; ultrasonic arrays; 20 s; 256 element phased array; MR guided in vivo porcine experiments; deep seated tissue treatment; in vivo liver tissue coagulation; kidney; large animal model; large scale phased array; liver; medical instrumentation; sonication; therapeutic ultrasound phased arrays; thigh; Animals; Coagulation; In vivo; Large-scale systems; Lesions; Liver; Phased arrays; Surgery; Thigh; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 1998. Proceedings., 1998 IEEE
  • Conference_Location
    Sendai
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-4095-7
  • Type

    conf

  • DOI
    10.1109/ULTSYM.1998.765241
  • Filename
    765241