DocumentCode :
2821253
Title :
Measurement of the regional exposure quality in quantitative coronary angiography
Author :
Wunderlich, W. ; Fischer, F. ; Agrawal, R. ; Arntz, H.R. ; Behrens, S. ; Schultheiss, HP ; Morguet, AJ
Author_Institution :
B. Franklin Univ. Hosp., Free Univ. at Berlin, Germany
fYear :
2000
fDate :
2000
Firstpage :
671
Lastpage :
674
Abstract :
This study determines the regional exposure quality and the impact of specific exposure types on size measurement errors in quantitative coronary angiography. The regional brightness distribution of the edge detection intensity matrix is calculated and presented as a histogram. Images of interventional devices were used to define the brightness distribution of 4 exposure types: (1) optimal quality (2) low contrast, (3) under- and (4) over-exposure. The size and brightness distribution of 254 imaged PTCA or stent carrier balloons were measured using QCA. The differences (in percent) between the measured and listed size were assigned to the identified exposure types. Optimal exposure (n=88) allowed the most valid measurements (accuracy±precision 1.1±3.3%). Low contrast (n=107) led to the least precise measurements (2.4±12.2%). Under-exposure (n=55) caused an evident size under-estimation (-12.2±9.20%). Overexposure (n=3) was under-represented in the study The regional brightness distribution provides an objective assessment of vessel exposure quality and predicts measurement errors
Keywords :
angiocardiography; diagnostic radiography; edge detection; measurement errors; size measurement; edge detection intensity matrix; low contrast; medical diagnostic imaging; optimal quality; overexposure; quantitative coronary angiography; regional brightness distribution; regional exposure quality measurement; specific exposure types impact; stent carrier balloons; underexposure; Angiography; Bismuth; Brightness; Cardiology; Histograms; Image edge detection; Image segmentation; Quantum cellular automata; Radiography; Size measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology 2000
Conference_Location :
Cambridge, MA
ISSN :
0276-6547
Print_ISBN :
0-7803-6557-7
Type :
conf
DOI :
10.1109/CIC.2000.898613
Filename :
898613
Link To Document :
بازگشت