• DocumentCode
    286253
  • Title

    In or out of specification?

  • Author

    Wotherspoon, N.

  • Author_Institution
    Hewlett Packard Ltd., Stockport
  • fYear
    1993
  • fDate
    34100
  • Firstpage
    42401
  • Lastpage
    42404
  • Abstract
    Modern quality systems require that after a suitable time interval (normally referred to as the calibration interval) an instrument is recalibrated, in order to verify that the product still performs to the published warranted specifications (i.e. to assure the continued accuracy of the product). The need to know the `actual´ measured value and the associated uncertainty is discussed and the problem of using the `significantly out of tolerance´ classification is considered
  • Keywords
    calibration; measurement errors; accuracy; calibration; instrument; measurement errors; specification; tolerance;
  • fLanguage
    English
  • Publisher
    iet
  • Conference_Titel
    Uncertainties in Electrical Measurement, IEE Colloquium on
  • Conference_Location
    London
  • Type

    conf

  • Filename
    243113