• DocumentCode
    2932350
  • Title

    Pre-reconstruction rigid body registration for positron emission tomography: An initial validation against ground truth

  • Author

    Nordberg, Peter ; Declerck, Jérôme ; Brady, Sir Michael

  • Author_Institution
    Dept. of Eng. Sci., Univ. of Oxford, Oxford, UK
  • fYear
    2010
  • fDate
    Aug. 31 2010-Sept. 4 2010
  • Firstpage
    5612
  • Lastpage
    5615
  • Abstract
    Our recent adaptation to PET of the method of Fitchard et al., for rigid body registration of CT sinograms enables motion between two temporal frames of PET data to be estimated and corrected prior to reconstruction. This avoids both the computation required by multiple reconstructions and the need to make choices regarding reconstruction methods that influence the images produced, and potentially change the estimated motion. Using realistic, simulated data with known ground truth, we report an initial investigation into the performance of the method as the number of counts and the accuracy with which the scan is divided into frames, corresponding to different positions, varies.
  • Keywords
    image reconstruction; image registration; medical image processing; motion estimation; positron emission tomography; CT sinograms; image reconstruction; motion estimation; positron emission tomography; prereconstruction rigid body registration; Accuracy; Attenuation; Biomedical imaging; Computed tomography; Data models; Image reconstruction; Positron emission tomography; Computer Simulation; Image Processing, Computer-Assisted; Positron-Emission Tomography; Reproducibility of Results; Rotation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2010 Annual International Conference of the IEEE
  • Conference_Location
    Buenos Aires
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-4123-5
  • Type

    conf

  • DOI
    10.1109/IEMBS.2010.5626800
  • Filename
    5626800