DocumentCode
2952297
Title
Test system for clinical force platforms
Author
Sanhueza, Germán Blanchard
Author_Institution
Sci. Fac., Univ. de Valparaiso, Valparaiso, Chile
fYear
2010
fDate
Aug. 31 2010-Sept. 4 2010
Firstpage
5772
Lastpage
5775
Abstract
Our team has built a system for testing and calibration of clinical Force Platforms, able to generate a circular path Center of Pressure (CoP) that can be accurately applied in the horizontal plane at a given travel speed, selectable radius and vertical force. The CoP generated, which is applied directly onto the platform, is used as a precise reference for both static and dynamic measurements, to check the operation of unknown Force Platforms performance characteristics or to verify specifications indicated by the Force Platform´s manufacturer. To confirm the feasibility and effectiveness of the proposed test apparatus, it was applied to a commercial clinical usage Force Platform manufactured by artOficio LTDA. The main advantages of the proposed Test System are the accuracy in fixing or setting a direct reference and the ability to maintain continuous monitoring through the generation of flexible and stable circular patterns during the tests.
Keywords
biomechanics; biomedical measurement; calibration; force; testing; circular path center of pressure; clinical force platforms; Accuracy; Dynamics; Force; Force measurement; Frequency measurement; Frequency modulation; Position measurement; Biomechanics; Fourier Analysis; Humans; Physical Medicine; Postural Balance; Pressure;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2010 Annual International Conference of the IEEE
Conference_Location
Buenos Aires
ISSN
1557-170X
Print_ISBN
978-1-4244-4123-5
Type
conf
DOI
10.1109/IEMBS.2010.5627840
Filename
5627840
Link To Document