    Magnetometer calibration set-up controlled by nuclear magnetic resonance

  • Author

    Weyand, K.

  • Author_Institution
    Phys. Tech. Bundesanstalt, Berlin, Germany
  • fYear
    1998
  • fDate
    6-10 July 1998
  • Firstpage
    580
  • Lastpage
    581
  • Abstract
    In order to calibrate magnetometer sensors in the field range from 2 mT to 150 mT a measuring setup consisting of a water-cooled multilayer solenoid and an NMR-magnetometer of the marginal oscillator type has been developed. With a coil constant of about 20 mT/A the field coil provides a field profile with two flat regions, where the field maxima are equal within 10/sup -5/ and with a distance of more than 40 mm so that both magnetometer readings can simultaneously be compared. The NMR-magnetometer operates with samples having an absorption line width of 2 /spl mu/T with a sample volume of 2 cm/sup 3/ for fields below 20 mT and of 1 cm/sup 3/ for higher fields.
  • Keywords
    calibration; magnetometers; nuclear magnetic resonance; solenoids; 2 to 150 mT; absorption line width; calibration set-up; coil constant; field coil; field maxima; field profile; flat regions; magnetometer calibration; magnetometer readings; marginal oscillator type; nuclear magnetic resonance; sample volume; water-cooled multilayer solenoid; Absorption; Calibration; Coils; Magnetic field measurement; Magnetic multilayers; Magnetic sensors; Magnetometers; Nuclear magnetic resonance; Oscillators; Packaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1998 Conference on
  • Conference_Location
    Washington, DC, USA
  • Print_ISBN
    0-7803-5018-9
  • Type

    conf

  • DOI
    10.1109/CPEM.1998.700065
  • Filename
    700065