Title :
High speed cell stiffness evaluation toward 100% reliability
Author :
Hirose, Yuki ; Kaneko, Makoto ; Kawahara, Tomohiro ; Yamanishi, Yoko ; Arai, Fumihito
Author_Institution :
Sch. of Mech. Eng., Osaka Univ., Suita, Japan
Abstract :
This paper proposes a high speed cell stiffness evaluation with high reliability. The procedure includes both online measurement and offline inspection. By using the property where the passing time of the cell in the micro channel is a function of the cell stiffness and its size, a real time vision system whose handling frequency is 6 kHz can capture the size as well as the passing time of each cell. During the online measurement, we save two image data at both the entrance and the exit for each cell. In offline manual inspection, all doubtful data are killed so that we can make sure the reliability. Experimental results ensure that the proposed approach can handle 500 cells within 10min in including both online and offline procedures.
Keywords :
microchannel flow; microsensors; reliability; doubtful data; high reliability; high speed cell stiffness evaluation; image data; microchannel; offline inspection; online measurement; real time vision system; Brightness; Cameras; Force; Red blood cells; Reliability; Size measurement; Time measurement;
Conference_Titel :
Sensors, 2011 IEEE
Conference_Location :
Limerick
Print_ISBN :
978-1-4244-9290-9
DOI :
10.1109/ICSENS.2011.6127399