Abstract :
We report on a capacitive micromachined ultrasonic transducer (CMUT) featuring isolation posts (PostCMUT) to solve a device reliability problem caused by charging during fabrication and operation, and to extend the device operation range beyond collapse voltage. The PostCMUTs were fabricated using a newly developed process based on the wafer-bonding technique. Paired tests showed the superior reliability characteristics of the PostCMUT design compared to those of earlier CMUT designs. The PostCMUTs showed no hysteresis during membrane-post contact.
Keywords :
capacitive sensors; dielectric hysteresis; micromechanical devices; piezoelectric transducers; reliability; ultrasonic transducers; wafer bonding; CMUT; capacitive charging; capacitive micromachined ultrasonic transducer; collapse voltage; device reliability problem; hysteresis; isolation posts; piezoelectric transducers; wafer-bonding technique; Biomembranes; Dielectrics; Electrodes; Fabrication; Hysteresis; Laboratories; Silicon; Testing; Ultrasonic transducers; Voltage;