DocumentCode :
3005346
Title :
Non-contact biopotential measurement from the human body using a low-impedance charge amplifier
Author :
Smith, Wayne J. ; LaCourse, John R.
Author_Institution :
Dept. of Electr. & Comput. Eng., New Hampshire Univ., Durham, NH, USA
fYear :
2004
fDate :
17-18 April 2004
Firstpage :
31
Lastpage :
32
Abstract :
This paper summarizes the preliminary design and tests of a non-contact biopotential electrode. Results of a non-contact ECG recording at 6 mm and 10 mm from the scalp are presented and compared to a body-surface ECG recording using wet electrodes. Some comments are made on its potential use as an EEG electrode.
Keywords :
bioelectric potentials; biomedical electrodes; electrocardiography; electroencephalography; EEG electrode; body-surface ECG recording; human body; low-impedance charge amplifier; noncontact biopotential electrode; scalp ECG recording; wet electrodes; Anthropometry; Bioelectric phenomena; Capacitance; Charge measurement; Current measurement; Electrocardiography; Electrodes; Feedback; Humans; Resistors;