Title :
A 2.4GHz non-contact biosensor system for continuous vital-signs monitoring
Author :
Ichapurapu, Ravi ; Jain, Suyash ; John, Gregory ; Monday, Timothy ; Lie, Donald Y C ; Banister, Ron ; Griswold, John
Author_Institution :
Dept. of Electr. & Comput. Eng., Texas Tech Univ., Lubbock, TX, USA
Abstract :
A non-contact vital-signs monitoring sensor system has been designed and tested. The system is assembled using two antennas instead of using one antenna and a circulator to achieve improved data accuracy with reduced DC offset. The system has been realized using a custom-designed analog baseband signal processing PCB that successfully replaces bulky baseband electronic blocks for system miniaturization. The sensor system was implemented in both single-channel and quadrature receiver configurations with the received signal being demodulated using several different algorithms. The vital-signs data extracted from a quadrature receiver with arctangent demodulation and autocorrelation method is found to be most accurate and reliable, achieving a mean error of +/- 1 beats/min vs. the reference heartbeat signal.
Keywords :
biosensors; patient monitoring; DC offset; antennas; continuous vital signs monitoring; frequency 2.4 GHz; noncontact biosensor; system miniaturization; Assembly systems; Baseband; Biomedical signal processing; Biosensors; Data mining; Demodulation; Monitoring; Sensor systems; Signal processing algorithms; System testing;
Conference_Titel :
Wireless and Microwave Technology Conference, 2009. WAMICON '09. IEEE 10th Annual
Conference_Location :
Clearwater, FL
Print_ISBN :
978-1-4244-4564-6
Electronic_ISBN :
978-1-4244-4565-3
DOI :
10.1109/WAMICON.2009.5207236