DocumentCode
3032167
Title
A total ionising dose, in-situ test campaign of DS18B20 temperature sensors
Author
Hofman, Jiri ; Sharp, Richard
Author_Institution
Isotron Ltd., Didcot, UK
fYear
2011
fDate
19-23 Sept. 2011
Firstpage
871
Lastpage
876
Abstract
Comprehensive tests are being performed to identify if the Dallas DS18B20 smart digital temperature sensor is (i) a potential solution for temperature monitoring during in-situ gamma irradiation tests and (ii) a suitable candidate for commercial low cost space applications. This work presents and discusses an applied test methodology for total ionising dose (TID) irradiation using a cobalt-60 gamma source. Preliminary test results are also published in this paper.
Keywords
intelligent sensors; radiation effects; temperature sensors; test equipment; Dallas DS18B20 smart digital temperature sensor; TID irradiation; cobalt-60 gamma source; in-situ gamma irradiation tests; in-situ test campaign; temperature monitoring; total ionising dose; Accuracy; EPROM; Measurement errors; Radiation effects; Sensitivity; Temperature measurement; Temperature sensors; automated test equipment; in-situ test; intelligent sensors; temperature sensors; total ionising dose;
fLanguage
English
Publisher
ieee
Conference_Titel
Radiation and Its Effects on Components and Systems (RADECS), 2011 12th European Conference on
Conference_Location
Sevilla
ISSN
0379-6566
Print_ISBN
978-1-4577-0585-4
Type
conf
DOI
10.1109/RADECS.2011.6131319
Filename
6131319
Link To Document