• DocumentCode
    3045302
  • Title

    Human erythrocyte sizing and deformability study by laser flow cytometer

  • Author

    Liu, Y. ; Fisher, A.C.

  • Author_Institution
    Sch. of Biomed. Eng., Univ. of Dundee, Dundee, UK
  • Volume
    1
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    324
  • Lastpage
    325
  • Abstract
    An erythrocyte size and deformability measurement method using flow cytometric technique is described in this paper. The instrument employed is a slit scanning based laser flow cytometer. The measurement is made of laser absorption. Native erythrocytes, rigid cells with different extent fixation, and calibration particles are studied. It has been found that the scanning pulse width has a good linear relationship with the size of the particles, scanning pulse height is however not linearly proportional to the particle size. It has also been found that the deformation of fresh erythrocytes in the focused flow system is insignificant.
  • Keywords
    biomedical measurement; calibration; cellular biophysics; laser applications in medicine; calibration; deformability measurement; erythrocyte size; human erythrocyte sizing; laser flow cytometer; scanning pulse height; slit scanning; Biomedical measurements; Blood; Laser beams; Measurement by laser beam; Particle measurements; Shape; Size measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5760986
  • Filename
    5760986