DocumentCode
306303
Title
An advanced transcutaneous energy transformer for a rechargeable cardiac pacemaker battery by using a resonant DC to DC converter
Author
Nishimura, Toshi Hiro ; Eguchi, Tetsuji ; Hirachi, Katsuya ; Saito, Masao ; Maejima, Yasushi
Author_Institution
Dept. of Electr. & Electron. Eng., Oita Univ., Japan
Volume
1
fYear
1996
fDate
5-10 Aug 1996
Firstpage
524
Abstract
An improvement method for a flat secondary transcutaneous transformer for a resonant power converter that transfers several watts of power across a large variable human skin gap is presented and built. The secondary transformer is piled, and the coupling coefficient, the conversion efficiency and the temperature rise are measured with an experimental set up. The coupling coefficient and the conversion efficiency are increased compared with conventional transformers. The temperature rise is decreased compared with conventional transformers
Keywords
DC-DC power convertors; battery chargers; cardiology; equivalent circuits; pacemakers; power transformer testing; power transformers; prosthetic power supplies; resonant power convertors; secondary cells; skin; temperature measurement; thermal analysis; transformer windings; cardiac pacemaker; conversion efficiency; coupling coefficient; flat secondary winding; performance testing; rechargeable battery; resonant DC/DC power converter; temperature rise; transcutaneous power transformer; variable human skin gap; Amorphous materials; Batteries; Copper; Couplings; DC-DC power converters; Pacemakers; Power supplies; Resonance; Skin; Transformer cores;
fLanguage
English
Publisher
ieee
Conference_Titel
Industrial Electronics, Control, and Instrumentation, 1996., Proceedings of the 1996 IEEE IECON 22nd International Conference on
Conference_Location
Taipei
Print_ISBN
0-7803-2775-6
Type
conf
DOI
10.1109/IECON.1996.571008
Filename
571008
Link To Document