• DocumentCode
    307482
  • Title

    Contiguous lesions by radiofrequency multielectrode ablation

  • Author

    Panescu, Dorin ; Fleischman, Sidney D. ; Whayne, James G. ; Swanson, David K.

  • Author_Institution
    EP Technol. Inc., Suunyvale, CA, USA
  • Volume
    1
  • fYear
    1995
  • fDate
    20-25 Sep 1995
  • Firstpage
    275
  • Abstract
    Radiofrequency catheter ablation can cure cases of atrial fibrillation. Multielectrode ablation catheters that can create long and contiguous lesions may benefit the physician by both increasing the predictability of lesion profiles and by reducing procedure time. This study presents an analysis of thermal profiles in tissue under a steerable 8F catheter capable of creating contiguous lesions. The catheter carried three 3-mm electrodes spaced 3.4 mm apart
  • Keywords
    bioelectric phenomena; biothermics; cardiology; finite element analysis; patient treatment; radiofrequency heating; temperature distribution; 3 mm; 3.4 mm; atrial fibrillation; contiguous lesions; electrodes; lesion profiles; procedure time; radiofrequency catheter ablation; radiofrequency multielectrode ablation; steerable 8F catheter; thermal profiles; tissue; Catheters; Electrodes; Finite element methods; In vitro; In vivo; Isothermal processes; Lesions; Radio frequency; Temperature distribution; Temperature sensors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1995., IEEE 17th Annual Conference
  • Conference_Location
    Montreal, Que.
  • Print_ISBN
    0-7803-2475-7
  • Type

    conf

  • DOI
    10.1109/IEMBS.1995.575107
  • Filename
    575107