Abstract :
Radiofrequency catheter ablation can cure cases of atrial fibrillation. Multielectrode ablation catheters that can create long and contiguous lesions may benefit the physician by both increasing the predictability of lesion profiles and by reducing procedure time. This study presents an analysis of thermal profiles in tissue under a steerable 8F catheter capable of creating contiguous lesions. The catheter carried three 3-mm electrodes spaced 3.4 mm apart
Keywords :
bioelectric phenomena; biothermics; cardiology; finite element analysis; patient treatment; radiofrequency heating; temperature distribution; 3 mm; 3.4 mm; atrial fibrillation; contiguous lesions; electrodes; lesion profiles; procedure time; radiofrequency catheter ablation; radiofrequency multielectrode ablation; steerable 8F catheter; thermal profiles; tissue; Catheters; Electrodes; Finite element methods; In vitro; In vivo; Isothermal processes; Lesions; Radio frequency; Temperature distribution; Temperature sensors;