    Effect of various metal areas with the same shadow area on defibrillation electrode resistance

  • Author

    Bonner, Matthew ; Mehra, Rahul

  • Author_Institution
    Medtronic Inc., Minneapolis, MN, USA
  • Volume
    1
  • fYear
    1995
  • fDate
    20-25 Sep 1995
  • Firstpage
    303
  • Abstract
    The objective of this paper is to determine if the electrode resistance for defibrillation electrodes with the same shadow area changes with the exposed metal area. To accomplish this the electrode resistance was compared for identical coil electrodes with and without insulation between the windings (backfill), and for solid cylinders of the same dimensions as the coil electrodes. The results were that removing the backfill reduced the resistance by 2%, compared with an area increase of 400%. Similarly, the resistance for the cylinder was 3% less than the resistance for the backfilled electrode even the though the surface area was 300% larger. Thus increasing the exposed metal area while keeping the same shadow area results in a small but significant decrease in resistance
  • Keywords
    defibrillators; backfilled electrode; cardiac arrhythmias treatment; coil electrodes; defibrillation electrode resistance; exposed metal area; shadow area; solid cylinders; Biomedical electrodes; Coils; Defibrillation; Electric shock; Immune system; Insulation; Solids; Surface resistance; Testing; Wire;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1995., IEEE 17th Annual Conference
  • Conference_Location
    Montreal, Que.
  • Print_ISBN
    0-7803-2475-7
  • Type

    conf

  • DOI
    10.1109/IEMBS.1995.575121
  • Filename
    575121