Effect of various metal areas with the same shadow area on defibrillation electrode resistance
Author
Bonner, Matthew ; Mehra, Rahul
Author_Institution
Medtronic Inc., Minneapolis, MN, USA
Volume
1
fYear
1995
fDate
20-25 Sep 1995
Firstpage
303
Abstract
The objective of this paper is to determine if the electrode resistance for defibrillation electrodes with the same shadow area changes with the exposed metal area. To accomplish this the electrode resistance was compared for identical coil electrodes with and without insulation between the windings (backfill), and for solid cylinders of the same dimensions as the coil electrodes. The results were that removing the backfill reduced the resistance by 2%, compared with an area increase of 400%. Similarly, the resistance for the cylinder was 3% less than the resistance for the backfilled electrode even the though the surface area was 300% larger. Thus increasing the exposed metal area while keeping the same shadow area results in a small but significant decrease in resistance
Keywords
defibrillators; backfilled electrode; cardiac arrhythmias treatment; coil electrodes; defibrillation electrode resistance; exposed metal area; shadow area; solid cylinders; Biomedical electrodes; Coils; Defibrillation; Electric shock; Immune system; Insulation; Solids; Surface resistance; Testing; Wire;