Abstract :
The infusion of contrast solution into multiple patients from a single package is desired in clinics concerned about the high cost of the solution. The use of check valves in the infusion line is a method of protection used to reduce the risk of blood-borne cross-contamination. Through testing of the operating characteristics of the valves and infusion system and the high back pressure testing of the valves, the suitability of this method is assessed. The results indicate that the use check valves reduce the risk of now from the patient to the injector. Results also indicate that the use of sprung valves, those which require a significant forward pressure to open them are more appropriate for this application
Keywords :
biomedical equipment; computerised tomography; diagnostic radiography; safety; valves; biosafety; blood-borne cross-contamination risk reduction; contrast dyes; contrast solution delivery; forward pressure; high back pressure testing; low osmolar contrast solutions; medical diagnostic imaging; medical instrumentation; multiple patients; one-way check valves integrity validation; sprung valves; Biomedical imaging; Circuit testing; Control systems; Hospitals; Medical diagnostic imaging; Mercury (metals); Packaging; Pressure control; Pumps; Valves;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1995., IEEE 17th Annual Conference
Conference_Location :
Montreal, Que.
Print_ISBN :
0-7803-2475-7
Type :
conf
DOI :
10.1109/IEMBS.1995.575306
Filename :
575306
Link To Document :
بازگشت