DocumentCode
307732
Title
Development of an integrated detector for NMR microscopy
Author
Stocker, J.E. ; Peck, T.L. ; Franke, S.J. ; Kruse, J. ; Feng, M. ; Magin, R.L.
Author_Institution
Dept. of Electr. & Comput. Eng., Illinois Univ., Urbana, IL, USA
Volume
1
fYear
1995
fDate
20-25 Sep 1995
Firstpage
843
Abstract
The development of an integrated detector consisting of a planar microcoil and low noise preamplifier for nuclear magnetic resonance (NMR) microscopy is described. The detector is designed to operate at 500 MHz (11.75 T). These customized detectors may be particularly important for NMR investigations of single cells, where it may become impractical to passively impedance match microcoils with extremely small values of impedance to 50 Ω or to electrically connect the microcoils to remotely located impedance matching and preamplifier circuitry
Keywords
NMR spectroscopy; UHF amplifiers; UHF detectors; biological NMR; biological techniques; coils; preamplifiers; 11.75 T; 500 MHz; MESFET preamplifier; NMR microscopy; customized detectors; impedance matching; integrated detector; low noise preamplifier; nuclear magnetic resonance; planar microcoil; preamplifier circuitry; single cells; Circuits; Coils; Detectors; Impedance matching; Microscopy; Nuclear magnetic resonance; Preamplifiers; Radio frequency; Signal to noise ratio; Spectroscopy;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1995., IEEE 17th Annual Conference
Conference_Location
Montreal, Que.
Print_ISBN
0-7803-2475-7
Type
conf
DOI
10.1109/IEMBS.1995.575391
Filename
575391
Link To Document