• DocumentCode
    3112409
  • Title

    A Wearable Reflectance Pulse Oximeter for Remote Physiological Monitoring

  • Author

    Mendelson, Y. ; Duckworth, R.J. ; Comtois, G.

  • Author_Institution
    Dept. of Biomed. Eng., Worcester Polytech. Inst., MA
  • fYear
    2006
  • fDate
    Aug. 30 2006-Sept. 3 2006
  • Firstpage
    912
  • Lastpage
    915
  • Abstract
    To save life, casualty care requires that trauma injuries are accurately and expeditiously assessed in the field. This paper describes the initial bench testing of a wireless wearable pulse oximeter developed based on a small forehead mounted sensor. The battery operated device employs a lightweight optical reflectance sensor and incorporates an annular photodetector to reduce power consumption. The system also has short range wireless communication capabilities to transfer arterial oxygen saturation (SpO2), heart rate (HR), body acceleration, and posture information to a PDA. It has the potential for use in combat casualty care, such as for remote triage, and by first responders, such as firefighters
  • Keywords
    blood vessels; cardiology; optical sensors; oximetry; oxygen; patient care; patient monitoring; annular photodetector; arterial oxygen saturation; battery operated device; bench testing; body acceleration; casualty care; forehead mounted sensor; heart rate; lightweight optical reflectance sensor; posture information; remote physiological monitoring; short range wireless communication; trauma injuries; wearable reflectance pulse oximeter; wireless pulse oximeter; Biomedical monitoring; Heart rate; Injuries; Optical pulses; Optical sensors; Reflectivity; Remote monitoring; Testing; Wearable sensors; Wireless sensor networks;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2006. EMBS '06. 28th Annual International Conference of the IEEE
  • Conference_Location
    New York, NY
  • ISSN
    1557-170X
  • Print_ISBN
    1-4244-0032-5
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2006.260137
  • Filename
    4461901