• DocumentCode
    3125818
  • Title

    Pressure microsensing catheters for neonatal care

  • Author

    Sauser, Frank E. ; Li, Chunyan ; Azizkhan, Richard G. ; Ahn, Chong H. ; Papautsky, Ian

  • Author_Institution
    Dept. of Electr. & Comput. Eng. & Comput. Sci., Cincinnati Univ., OH, USA
  • fYear
    2004
  • fDate
    24-27 Oct. 2004
  • Firstpage
    1476
  • Abstract
    A pressure microsensing catheter for neonatal care has been developed to measure intravascular blood pressure. The design uses polymer flip-chip bonding on a flexible Kapton film for mounting silicon pressure microsensors into the 5 French dual lumen neonatal catheters (1.67 mm o.d.). While other groups have mounted MEMS sensors on guide wires, our dual-lumen approach permits one lumen of the catheter to contain a packaged pressure sensor while the other lumen retains the catheter functionality. The packaged sensors were characterized in nitrogen and water environments, both statically and dynamically.
  • Keywords
    biosensors; blood pressure measurement; catheters; flip-chip devices; microsensors; paediatrics; polymers; pressure sensors; silicon; French dual lumen neonatal catheters; MEMS sensors; flexible Kapton film; intravascular blood pressure measurement; neonatal care; nitrogen; packaged pressure sensor; polymer flip-chip bonding; pressure microsensing catheter; silicon pressure microsensors; water; Blood pressure; Bonding; Catheters; Packaging; Pediatrics; Polymer films; Pressure measurement; Semiconductor films; Sensor phenomena and characterization; Silicon;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors, 2004. Proceedings of IEEE
  • Print_ISBN
    0-7803-8692-2
  • Type

    conf

  • DOI
    10.1109/ICSENS.2004.1426466
  • Filename
    1426466